Testing procedure – RADWAG WTB 600/AU Carat Balance User Manual
Page 40

Density of metal alloy [g/cm
3
]
Gold
caratage
[K]
% content
Copper
and silver
Copper Silver
Range of sample
density [g/cm
3
]
24
100 / 100
19.32
19.32
19.32
19.13 ~ 19.51
22
91.6 / 100
17.73
17.63
18.06
17.45 ~ 18.24
20
83.4 / 100
16.42
16.19
16.94
16.03 ~ 17.11
18
75.0 / 100
15.24
14.99
15.96
14.84 ~ 16.12
14
58.4 / 100
13.38
13.04
14.30
12.91 ~ 14.44
10
41.7 / 100
11.91
11.54
12.96
11.42 ~ 13.09
16.7.3. Testing procedure
• Select working mode
>
in accordance with point 16.1 of this
user manual, after which the main window of selected working mode is
displayed:
• If measurement result is other than zero press
key (balance
zeroing).
• Load tested sample on the top weighing pan (weighing in the air,
flickering command IN AIR) and on stabilization of measurement
result press
key,
• Simultaneously the following command is displayed:
• Unload the tested sample from the top weighing pan,
- 40 -