beautypg.com

Testing procedure – RADWAG WTB 600/AU Carat Balance User Manual

Page 40

background image

Density of metal alloy [g/cm

3

]

Gold

caratage

[K]

% content

Copper

and silver

Copper Silver

Range of sample

density [g/cm

3

]

24

100 / 100

19.32

19.32

19.32

19.13 ~ 19.51

22

91.6 / 100

17.73

17.63

18.06

17.45 ~ 18.24

20

83.4 / 100

16.42

16.19

16.94

16.03 ~ 17.11

18

75.0 / 100

15.24

14.99

15.96

14.84 ~ 16.12

14

58.4 / 100

13.38

13.04

14.30

12.91 ~ 14.44

10

41.7 / 100

11.91

11.54

12.96

11.42 ~ 13.09



16.7.3. Testing procedure

• Select working mode

>

in accordance with point 16.1 of this

user manual, after which the main window of selected working mode is
displayed:

• If measurement result is other than zero press

key (balance

zeroing).

• Load tested sample on the top weighing pan (weighing in the air,

flickering command IN AIR) and on stabilization of measurement

result press

key,

• Simultaneously the following command is displayed:

• Unload the tested sample from the top weighing pan,