beautypg.com

Zeroing, Weighing, Testing carat content of pure gold – RADWAG WTB 600/AU Carat Balance User Manual

Page 37

background image

16.4.3. Zeroing

To ZERO the indication press:

key.

The balance’s display shows mass indication equal to zero, and pictograms

and

appear.

Zeroing is an activity equal to setting a new zero point, comprehended by
the balance as the precise zero point. Zeroing is accessible only with stable
display status.

Caution:
Zeroing of display indication is possible only within the ±2% range from
balance’s maximal capacity.
If zeroed value is more than the ±2% of balance’s maximal capacity, the
display indicates message and short audible signal is emitted.



16.5. Weighing

It is the balance’s standard working mode (marked as: mode B)
on the display, which enables carrying out regular weighing processes.

If the measurement result is other than zero – press

key (balance

zeroing).

Place weighed object on balance’s weighing pan. As stability marker

is displayed, read measurement result.

At the same time the weighing measurement can be printed on a

connected printer by pressing button

.



16.6. Testing carat content of pure gold

Working mode MODE G enables testing pure gold for its caratage from
9.0K (37.5% of gold content) to 24.0K (100% of gold content).

Procedure:

• Select working mode

>

in accordance with point 16.1 of this

user manual. The main window of the working mode is displayed:

- 37 -