Send measurement result in basic measuring unit, Immediately, Eighing unit – RADWAG WTB 600/AU Carat Balance User Manual
Page 57: Send stable measurement result in current w, Frame format, Accessible responses
21.4.5. Immediately send measurement result in basic measuring unit
Format: SI CR LF
Accessible responses:
SI_I CR LF
- command understood but not accessible at this moment
MASS FRAME
- response is immediate with mass value in basic weighing unit
Frame format:
1 2
3
4
5
6 7-15
16 17 18 19 20 21
S I space
stability
marker
space sign mass space
unit
CR LF
Example:
S I CR LF – command sent from a computer
S I _ ? _ _ _ _ _ _ _ 1 8 . 5 _ k g _ CR LF - command carried out,
immediate response of mass value in basic weighing unit
21.4.6. Send stable measurement result in current weighing unit
Format: SU CR LF
Accessible responses:
SU_A CR LF
SU_E CR LF
- command understood and in progress
- time limit exceeded while waiting for stable measurement result
SU_I CR LF
- command understood but not accessible at this moment
SU_A CR LF
MASS FRAME
- command understood and in progress
- response is mass value in current measuring unit
Frame format:
1 2
3
4
5
6 7-15 16 17 18 19 20 21
S U space
stability
marker
space sign mass space
unit
CR LF
- 57 -