Volume Method
S E T S T R O K E T O 5 0 %
E N T E R T O S T A R T
1. Manually adjust the stroke length of the pump to 50%
2. ** Fill your calibration column to the zero or starting point
3. Press the ENTER key
T I M E R
X X S E C
E N T E R T O S T O P
4. Allow the pump to run for any length of time you choose, dependent on flow
rate, calibration column size, process conditions, etc. The system will time out
at a maximum of 300 seconds.
5. Press the ENTER key to stop the pump
E N T E R V O L U M E
V O L =
X X . X X X
6. Use the UP and DOWN arrow keys and the ENTER key to input the volume
pumped from the calibration column (in gallons or liters, dependent on MPC
setup).
7. Press ENTER to proceed
S E T S T R O K E T O 1 0 0 %
E N T E R T O S T A R T
8. Manually adjust the stroke length of the pump to 100%
9. ** Fill your calibration column to the zero or starting point
10. Press the ENTER key
T I M E R
X X S E C
E N T E R T O S T O P
11. Allow the pump to run for any length of time you choose, dependent on flow
rate, calibration column size, process conditions, etc. The system will time out
at a maximum of 300 seconds.
12. Press the ENTER key to stop the pump