beautypg.com

Campbell Scientific AVW200-series Vibrating Wire Interfaces User Manual

Page 84

background image

Appendix E. Status Table

Status

Fieldname

Description

User can change?

Rf_ForceOn

When Rf_ForceOn is set to 1 the radio is always on

ignoring the duty cycle setting.

Yes

Rf_Protocol

Identifies the radio protocol that will be used. The

AVW200 is always fixed at 2 (PakBus Aware mode)

Yes (changing this parameter to a

value of 1 will mess up the RF

communication). All other values will

revert to a value of 2.

RfSignalLevel

The signal level of every 5

th

PakBus packet received

over RF

Yes (clear to zero)

RfRxPakBusCnt Number of PakBus packets that have been received

over RF communication

Yes (clear to zero)

RfPwrMode

Radio power modes:
NO_RF (No Radio)
RF_ON (<24ma Always On)
RF_1/2_Sec (<4ma ½ Second)
RF_1_Sec (<2ma 1 Second)
RF_8_Sec (<0.4ma 8 Second)
RF_OFF (Radio Off)

Yes (ie to change from ½ seconds

duty cycle to 1 seconds duty cycle

mode, edit the parameter with:

RF_1_Sec

PortStatus(1)

Indicates control port 1 level
0 = off (low zero volts)
-1 = on (high five volts)

No

PortStatus(2)

Indicates control port 2 level
0 = off (low zero volts)
-1 = on (high five volts)

No

PortStatus(3)

Indicates control port 3 level
0 = off (low zero volts)
-1 = on (high five volts)

No

PortConfig(1)

Indicates control port 1 configuration (function

disabled reserved for future use).

Yes

PortConfig(2)

Indicates control port 1 configuration (function

disabled reserved for future use).

Yes

PortConfig(3)

Indicates control port 1 configuration (function

disabled reserved for future use).

Yes

MSPversion(1)

MSP430 CPU #1 OS version

No

MSPversion(2)

MSP430 CPU #2 OS version

No

MSPversion(3)

MSP430 CPU #3 OS version

No

MSPversion(4)

MSP430 CPU #4 OS version

No

MSPversion(5)

MSP430 CPU #5 OS version

No

MSPClkFreq(1)

MSP430 CPU #1 RC oscillator frequecy in Hz

No

MSPClkFreq(2)

MSP430 CPU #2 RC oscillator frequecy in Hz

No

E-2