2 wireless connection, Wireless connection – Campbell Scientific AVW200-series Vibrating Wire Interfaces User Manual

Page 20

background image

AVW200-series 2-Channel Vibrating Wire Spectrum Analyzer Modules

Check the manufacturer’s specification for the sensors frequency

and excitation range before picking the begin/end frequencies and

excitation voltage.

For example, the following AVW200() instructions can be used to

measure two sensors:

AVW200(Result,Com1,200,200,Dst(1,1),1,1,1,1000,3500,2,_60HZ,1,0)

AVW200(Result,Com1,200,200,Dst(2,1),2,1,1,1000,3500,2,_60HZ,1,0)

Where,

AVW200 connects to datalogger control ports 1 & 2 via 17855 cable

(option Com1)

Begin Frequency = 1000

End frequency = 3500

Excitation voltage = 12 V peak to peak (option 2)

Read more! A thorough description of the AVW200() instruction and its

parameters is provided in Section 6.1. A complete example program that

measures two vibrating wire sensors (no multiplexer) is included in

Section 7.1.1.

3.1.2 Wireless Connection

Cable that Comes with Sensor

19246 Power Cable

SC12 CS I/O Cable

Wireless Connection

Sensors

AVW206

Sensors

Power

Supply

Datalogger

RF401

Onsite

Office

For this example configuration, the sensor(s) are attached directly to the

AVW206. The AVW206 interface transmits the data to an RF401 spread

spectrum radio that is connected to the datalogger.

At the AVW206 site, do the following steps:

1. Configure the AVW206 for RF communications as described in Section

5.1, Connecting to DevConfig, and Section 5.2.1, Communications.

2. Attach the vibrating wire sensor(s) to the AVW206 as shown in FIGURE

4-1.

NOTE