AVW200-series 2-Channel Vibrating Wire Spectrum Analyzer Modules


When using the “Mcmm” terminal command, no other

method of measurement should be used or multiplexing will

get out of sequence and measurement errors will result.

Read more! Appendix D lists the fields in the public table and provides a

brief description of each. Appendix E provides a comprehensive list of

status table variables with brief descriptions.

6. Programming

6.1 AVW200 Instruction

If using SDI-12 to communicate with the AVW200, then use the

SDI12Recorder instruction to trigger and retrieve measurements

from the AVW200 (see Section 6.2, SDI-12 Measurements.)

When using the RS-232 (PakBus) or RF(PakBus) protocol, the datalogger is

programmed using the AVW200 instruction in CRBasic.

Please ensure that the AVW200 instruction is included in

LoggerNet. If using LoggerNet 3.4.1 or lower, the user needs to

download the most recent version of the operating system (OS) for

the datalogger. This installation installs the required CRBasic

files on the user’s computer so that the AVW200 instruction

shows up in the editor.

The AVW200 instruction is used to read measurements from one or more

vibrating wire sensors when the sensors are connected to the AVW200

channels directly or connected to a multiplexer that’s attached to the AVW200.

In addition to frequency, diagnostic information is returned for the sensor (see

the Dest parameter).

The datalogger program can run the AVW200 instruction in either the pipeline

mode (Section 6.1.1, Pipeline Mode) or sequential mode (Section 6.1.2,

Sequential Mode). In the pipeline mode, the first execution of the instruction

sets up the AVW200; subsequent execution intervals retrieve the data values. If

different beginning and ending frequencies are required to measure different

types of sensors, use multiple AVW200 instructions with different beginning

and ending frequencies specified in each instruction. The sequential mode

performs each instruction in sequence; waits for each instruction completion;

and then repeats this process for each execution interval. The minimum scan

rate for an AVW200 program is 2 seconds per sensor.

Syntax

AVW200 (Result, ComPort, NeighborAddr, PakBusAddr, Dest, AVWChan,

MuxChan, Reps, BeginFreq, EndFreq, ExVolt, Therm50_60Hz, Multiplier,

Offset, [Optional] AmpThreshold)

Result

The Result parameter is a variable that indicates the success

or failure of the datalogger's communication attempt with

the AVW200. If more than one AVW200 is used and the

CAUTION

NOTE

NOTE

37