Appendix D. The Public Table

The public table of the AVW200 displays the current sensor measurement

values as well as the current settings (see TABLE D-1).
When the DeviceConfig runs the troubleshooter, it forces a measurement by

writing to the Timeseries() array in the Public table. When the

Timeseries(1)..Timeseries(4) elements are written with the proper values a

measurement is performed and the files Timeseries.bin and Spectrum.bin are

created or over-written if previous measurements have been forced. These file

are then used by DeviceConfig to graph the time domain and frequency domain

graphs in the troubleshooter.
A Datalogger program can also force a measurement by using the

GetVariable() and/or SetVariable() instructions (see Program Example D.1).

To do this, use the CRBasic SetVariable() instruction to set the TimeSeries(2),

TimeSeries(3) and TimeSeries(4) variables. Once these variables have the

correct values for Begin, End Frequency and Excite voltage, the measurement

can be forced by writing the TimeSeries(1) with the AVW200 channel and

Mux Channel (ie 101 = AVW200 Chan1 and MuxChan1, 205 = AVW200

Chan1 and MuxChan5, or 208 = AVW200 Chan2 and MuxChan8). The

measurement is completed when the TimeSeries(1) value is zeroed by the

AVW200. Once zeroed the TimeSeries(5) through TimeSeries(11) values can

be read using the CRBasic GetVariable() instruction.

TimeSeries(5) through TimeSeries(11) were added to OS version

STD.02. Customers with OS version STD.01 can get these public

variables by downloading a new OS (refer to Section 5.4, Send

OS).

TABLE D-1. Description of the Public Table

Record No Current record number for this set of data
Time

Stamp

Time the record was recorded

Control Parameters for AVW200 Instruction Communications

Result

Result of the last measurement

NumReps

Number of replications specified for this channel

Chan

Channel this record was recorded from

MuxChan

Multiplexer channel this data was recorded from

BeginFreq

The begin frequency used for this measurement

EndFreq

The end frequency used for this measurement

ExciteVolts The excitation voltage used for this measurement
Do_Therm Do the thermistor measurement or not
RScnt

Count down to start of measurement for the RS-232

communication port. -1 if no interval or not used

RFcnt

Count down to start of measurement for the RF

communication port. -1 if no interval or not used

NOTE

D-1