AVW200-series 2-Channel Vibrating Wire Spectrum Analyzer Modules

Read more! You can find an example program that converts resistance to

temperature in Section 7.1.2, and detailed information about the

thermistors in Appendix B.

3. Quick Start Guides

The AVW200 can be used in many types of systems—from simple to complex.

The following quick start guides provide steps used to set up a system for some

example configurations.

3.1 One or Two Sensors (no multiplexers)

3.1.1 Direct RS-232 Connection

Cable that Comes with Sensor

17855 Pigtailed Cable or

18663 Null Modem Cable

19246 Power Cable

Sensors

AVW200

Sensors

Datalogger

Power

Supply

For this simple configuration, the AVW200 can be used straight from the box

(i.e., settings do not need to be changed in DevConfig). The sensor(s) are

attached directly to the AVW200. The AVW200 is connected directly with the

datalogger via the 17855 cable or 18663 cable. The 17855 cable terminates in

pigtails for connection to datalogger control port pairs (C1/C2…C7/C8). The

18663 Null Modem Cable has a DB9 connector for attachment to the

datalogger’s RS-232 port.

The following steps are used to measure the sensor(s):

1. Attach the vibrating wire sensor(s) to the AVW200 as shown in

FIGURE 4-1.

2. Use the 17855 cable to attach the AVW200 to a control port pair on the

datalogger (i.e., C1/C2, C3/C4, C5/C6, C7/C8), or use the 18663 Null

Modem cable to attach the AVW200 to the RS-232 port on the datalogger.

3. Connect one end of the 19246 power cable to the 12V and G terminals on

the AVW200 and the other end to the 12V and G terminals on the

datalogger or external power supply.

Read more! You can find power connection information in Section 4.2,

and datalogger connection information in Section 4.3.

4. Create a CRBasic program that includes an AVW200() instruction for each

of the sensors.