User's Guide _______________________________________________________________________

184 __________________________________________________________________ M210762EN-G

Device Identification Objects

DMT345/346 Modbus conforms to the extended identification level
defined in the Modbus Application Protocol Specification V1.1b. Both
stream access and individual access to the objects is supported.

Table 54

Modbus Device Identification

Object Id

Object Name

Description

0x00

VendorName

“Vaisala”

0x01

ProductCode

Product code (e.g. “DMT340”)

0x02

MajorMinorVersion

Software version (e.g. “5.10”)

0x03

VendorUrl

“http://www.vaisala.com/”

0x04

ProductName

Product name of the instrument

0x80

SerialNumber

Serial number (e.g. “D0610020”)

0x81

CalibrationDate

Date of the last calibration
(e.g. “2011-03-01”, empty if not
available)

0x82

CalibrationText

Information text of the last calibration
(empty if not available)

Exception Responses

Exception responses from the transmitter are according to the Modbus
Application Protocol Specification V1.1b.

Table 55

Modbus Exception Responses

Code

Name

Reason

01

ILLEGAL FUNCTION

Unsupported function code

02

ILLEGAL DATA ADDRESS

Address out of valid ranges

03

ILLEGAL DATA VALUE

Otherwise invalid request


Accessing unavailable (unsupported or temporarily missing) registers
inside ranges specified in Table 48 does not generate an exception.
“Unavailable” value (a quiet NaN for floating point data or zero for
integer data) is returned instead. An exception is generated only for any
access outside the register blocks defined in section Register Map on
page 178.