beautypg.com

Opening and closing adjustment mode, Table 34, Indicator led functions – Vaisala DMT346 User Manual

Page 150

background image

User's Guide _______________________________________________________________________

150 __________________________________________________________________ M210762EN-G

Opening and Closing Adjustment Mode

Open the transmitter cover. The buttons needed in the adjustment are on
the left-hand side of the motherboard, see Figure 2 on page 23.

Press the ADJ button to enable the adjustment mode. The indicator LED
indicates the adjustment availability.

Table 34

Indicator Led Functions

Indicator Led Function

Description

LED off

adjustment locked

LED on

adjustment available

LED blinking evenly

measurement not stabilized

LED blinking with short pulses

performing Purge/AutoCal

Press the ADJ button again to disable the adjustment mode.

NOTE

Fixed pressure compensation of 1013.25 hPa is used when in adjustment
mode. Adjustments should be performed at ambient pressures.