Page 101

background image

Chapter 4 _________________________________________________________________ Operation

VAISALA _______________________________________________________________________ 101

NOTE

Pressure compensation is intended to be used in normal air only. When
measuring in other gases, please contact Vaisala for further information.

NOTE

Fixed pressure compensation value of 1013.25 hPa is used when in
adjustment mode.

Using Display/Keypad

Use display/keypad to set the pressure compensation. To select the
pressure unit using display/keypad see section Changing Quantities and
Units on pa
ge 97.

1.

Open the Main Menu

by pressing any of the ▲▼◄► arrow

buttons.

2.

Select Measuring

and press the ►arrow button to confirm your

selection.

3.

Select Pressure compensation

and press the ► arrow button to

confirm you selection.

4.

Press SET and enter the pressure value by using the arrow buttons.

5.

Press OK and EXIT to return to the basic display.

Using Buttons on Motherboard

Pressure set buttons (P

chk

and P

set

) can be used to set the process

pressure.

0504-052

Figure 58

Pressure Set Buttons on Motherboard

The following numbers refer to Figure 58 above:

1

=

Pressure set button

2

=

Pressure check button

Press check button (P

chk

) and a red LED will flash the current pressure

setting in bar

a

.