Data Filtering

The averaging data filter calculates a average over a certain period of
time. The lowest measurement noise is achieved with the extended
filtering. There are three filtering levels available.

Table 29

Filtering Levels

Setting

Filtering Level

OFF

No filtering.

STANDARD

Standard filtering, enabled by default.
Approximately 13 s moving average.

EXTENDED

Extended filtering. Approximately 1 min average by
default, but configurable on the serial line.


Use display/keypad to set the filtering level.

1.

Open the Main Menu

by pressing any of the ▲▼◄► arrow

buttons.

2.

Select Measuring

by pressing the ► arrow button.

3.

Select Filtering and press CHANGE to confirm your selection.

4.

Select Off/Standard/Extended and press SELECT to confirm
your selection.

5.

Press EXIT to return to the basic display.