Two-Point Relative Humidity Adjustment

Using Display/Keypad

NOTE

Reference humidities must be ~0 %RH and 30 ... 75 %RH. A suitable
product for the two point RH calibration is for example the Vaisala
Humidity Calibrator HMK15.

1.

Press the ADJ button (see Figure 2 on page 23) to open the
ADJUSTMENT MENU:

2.

Select Adjust T

d

measurement,

press ► button.

3.

Select 2-point RH adjustment, press START. At this point, the
device carries out the chemical purge.

4.

Remove the filter from the probe and insert the probe in the dry end
reference condition (~0 %RH). Press OK.

0706-018A

Figure 73

Inserting the Sensor in Reference Humidity 1

5.

Wait at least 30 minutes for the sensor to stabilize. Follow the
stabilization from the GRAPH display. Press EXIT to return to the
previous display.

6.

Press READY when stabilized. Enter the reference value by using
the arrow keys.

0706-019B

Figure 74

Waiting for Readings to Stabilize

Now proceed to the adjustment at the wet end reference condition
(30 ... 75 %RH) and carry out the procedure as described in the
previous items.