Using serial line, Xpres and pres, Table 28 – Vaisala DMT346 User Manual

Page 102: Conversion factors for pressure units

background image

User's Guide _______________________________________________________________________

102 __________________________________________________________________ M210762EN-G

Press the set button (P

set

) to set the pressure value. The number of presses

equals the number of bar

a

to be set (for example,

three presses = three bar

a

). After a few seconds, the red LED will confirm

the setting by flashing the new pressure value.

Using Serial Line

XPRES and PRES

Command XPRES must be used if the value is changed frequently (for
example, by an automatic system that updates the value). Its value is not
retained at reset, and when set to 0; last value set with PRES is used
instead. Use the serial line and do the following:

PRES [aaaa.a]

XPRES [aaaa.a]

where

aaaa.a = Absolute process pressure (hPa)

Examples:

>pres
Pressure : 1013.00 hPa ?
>pres 1010
Pressure : 1010.00 hPa
>

Table 28

Conversion Factors for Pressure Units

From

To: hPa

mbar

1

PaN/m2

0.01

mmHg torr

1.333224

inHg

33.86388

mmH

2

O

0.09806650

inH

2

O

2.490889

atm

1013.25

at

980.665

bar

1000

psi

a

1)

68.94757

1) psi

a

= psi absolute

Example:

29.9213 inHg = 29.9213 × 33.86388 hPa = 1013.25 hPa