Technical data, Specifications, Performance – Vaisala DMT346 User Manual
Page 161: Chapter 8, Table 35, Dewpoint specifications dmt345, Chapter 8, technical data, pr

Chapter 8 _____________________________________________________________ Technical Data
VAISALA _______________________________________________________________________ 161
CHAPTER 8
TECHNICAL DATA
This chapter provides the technical data of the product.
Specifications
Performance
Table 35
Dewpoint Specifications DMT345
Property
Value
Sensor
Vaisala DRYCAP®180S
Measurement range
-40 ... +100 °C (-40 … +212 °F) T
d
Accuracy
±2 °C (±3.6 °F) T
d
See accuracy graph in Figure 78 on
page 161.
Response time 63% [90%] flow rate
1l/min and 1 bar pressure
from dry to wet
5s [10 s]
from wet to dry including auto-
calibration
45s [5 min]
0605-043
Figure 78
Dewpoint Measurement Accuracy Graph DMT345