Prh p, Pp p ppm – Vaisala DMT346 User Manual
Page 175

Appendix B _______________________________________________________ Calculation Formulas
VAISALA _______________________________________________________________________ 175
The water vapor pressure is calculated using:
100
ws
w
P
RH
P
⋅
=
(7)
Parts per million by volume is calculated using:
(
)
w
w
v
P
p
P
ppm
−
⋅
=
6
10
(8)
Symbols:
T
d
=
dewpoint temperature (°C)
P
w
=
water vapour pressure (hPa)
P
ws
=
water vapour saturation pressure (Pa)
RH
=
relative humidity (%)
x
=
mixing ratio (g/kg)
p
=
atmospheric pressure (hPa)
a
=
absolute humidity (g/m
3
)
T
=
temperature (K)
h
=
enthalpy (kJ/kg)
This manual is related to the following products: