beautypg.com

Rsel – Vaisala DMT346 User Manual

Page 128

background image

User's Guide _______________________________________________________________________

128 __________________________________________________________________ M210762EN-G

0706-029

Figure 63

Relay Availability

Arrow in Figure 63 above shows where enabled relays are listed on the
display. Activation state shown in black. Disabled relays are not shown.

Use the display/keypad to set the relay outputs.

1.

Open the Main Menu

by pressing any of the ▲▼◄► arrow

buttons.

2.

Select Interfaces

, confirm by pressing the ►arrow button.

3.

Select Relay outputs

, confirm by pressing the ►arrow button.

4.

Select Relay 1/2/3/4

, confirm by pressing the ►arrow button.

5.

Select the Quantity, confirm by pressing Change. Select the
Quantity by using the arrow buttons. Confirm your selection by
pressing Select. (Press Fault Status when the relay follows the
transmitter error.) Press Change to set the value.

6.

Select Act. above / Act. below. Press SET to confirm your
selection. If asked, select MODIFY if you want to set the setpoint
by using the arrow buttons. Select REMOVE if you want to
remove the setpoint.

7.

Select Hysteresis. Press SET. Set the hysteresis by using the arrow
buttons. Press OK.

8.

Select Relay enable, press ON/OFF to enable/disable the relay.

RSEL

Use the serial line to select the quantity, setpoints and hysteresis or
enable/disable the relay outputs. Issue the RSEL command.

RSEL [q1 q2 q3 q4]

where

q1

=

Quantity for the relay 1 or Fault/Online

q2

=

Quantity for the relay 2 or Fault/Online

q3

=

Quantity for the relay 3 or Fault/Online

q4

=

Quantity for the relay 4 or Fault/Online