User's Guide _______________________________________________________________________

182 __________________________________________________________________ M210762EN-G

Configuration flags are used to select some basic options of the
instrument and to manually start the sensor operations.

Table 51

Configuration Flag Registers

Name

Address

Description

Standard filtering on/off

1281

1 = Filtering on

Extended filtering on/off

1282

1 = Extended filtering on

Automatic sensor purge on/off

1283

1 = Automatic purge on.
See configuration register
0773…0774 (1027).

Start-up sensor purge on/off

1284

1 = Start-up purge on

Sensor purge in progress

1285

1 = Purge in progress
(write 1 for manual start)

AutoCal in progress

1286

1 = AutoCal in progress
(write 1 for manual start)


Flags 1281 and 1282 are mutually exclusive.

Some of the configuration registers might not have an effect depending
on the instrument model and configuration.

Other configuration settings are not usually needed and are not available
using the Modbus interface. The service cable must be used to change
other configuration settings if needed.

Exception Status Outputs

Exception status outputs (read with function code 07) tell the summary of
the transmitter status as described in Table 52 below.

Table 52

Exception Status Outputs

Output

Name

Description

0 (0x01)

Fault status

1 = No errors

1 (0x02)

Online status

1 = Online data available

Status information is available also with register access; see section
Status Registers (Read-Only) on page 181.