Chapter 7 _____________________________________________________________ Technical Data

VAISALA ________________________________________________________________________ 71

CHAPTER 7

TECHNICAL DATA

This chapter provides the technical data of the HMT120.

Table 9

Relative Humidity Measurement Specifications

Property

Description / Value

Measurement range

0 ... 100 %RH

Accuracy (including non-linearity,

hysteresis, and repeatability):

at 0 ... +40 °C (+32 ... +104 °F)

±1.5 %RH (0 ... 90 %RH)

±2.5 %RH (90 ... 100 %RH)

at -40 ... 0 °C and +40 ... +80 °C

(-40 ... +32 °F and +104...+176 °F)

±3.0 %RH (0 ... 90 %RH)

±4.0 %RH (90 ... 100 %RH)

Factory calibration uncertainty at

+20 °C (+68 °F)

±1.5 %RH

Humidity sensor

HUMICAP

®

180R

Table 10

Temperature Measurement Specifications

Property

Description / Value

Measurement range

-40 ... +80 °C (-40 ... +176 °F)

Accuracy over temperature range:

at +15 ... +25 °C (+59 ... +77 °F)

±0.2 °C (±0.36 °F)

at 0 …+15 °C and +25 …+40 °C

(+32 ... +59 °F and +77 ...+104 °F) ±0.25 °C (±0.45 °F)

at -40 ... 0 °C and +40 ... +80 °C

(-40 ... +32 °F and +104...+176 °F) ±0.4 °C (±0.72 °F)

Temperature sensor

Pt1000 RTD Class F0.1 IEC 60751