Adjustment with hm70, Figure 17 – Vaisala HMT120 User Manual

Page 56

background image

User's Guide _______________________________________________________________________

54 ___________________________________________________________________ M211244EN-B

Adjustment with HM70

You can check and adjust the HMT120 relative humidity measurement

with the HM70 Hand-Held Humidity and Temperature meter. A HM70-

connection cable is needed. Vaisala order code: 211339.

There are four types of adjustments available: field checking and

adjustment using a calibrated reference probe, one-point adjustment

using a calibrator, two-point adjustment using a calibrator, and LiCl-

NaCl adjustment.

Follow the first seven steps and continue according to the chosen

adjustment method.

0505-351

Figure 17

Location of the MI70 Probe and Cable Connector

Ports

1.

Connect the 211339 HM70-connection cable to the

SERVICE PORT

connector on the HMT120 motherboard (see Figure 11 on page

21).

2.

Connect the other end of the connection cable to either of the

HM70's connector ports located on the bottom of the indicator (see

Figure 17 above).

3.

Turn on both devices (or just HM70, in case that HMT120 is on

continuously).

4.

Check that the transmitter reading is shown on the indicator. The

reading of the transmitter is shown on the top or middle row of the

display, depending on which connector port the connection cable is

connected to.

5.

Press the

ADJ

button on the HMT120 motherboard to open the

adjustment mode. Both LEDs on HMT120 motherboard remain in

OFF state and text "Starting adjustment mode for HMP110" is

shown at the MI70 display.

NOTE

In case of using a HMT120 transmitter with display option, text "MI70

adjustment mode" is shown on the upper row of HMT120 display.