Chapter 3 ________________________________________________________________ Installation

VAISALA ________________________________________________________________________ 19

Drill the holes for the duct installation kit as follows:

1.

Use a 24-mm drill bit to drill a hole to the duct wall for the

humidity probe.

2.

Drill holes for the duct installation kit mounting screws around the

hole in a square arrangement, 42 mm apart from each other. Use a

3.2-mm drill bit to drill the holes for the mounting screws (four

ST4.2×16-C-Z DIN 7981 screws).

Probe Mounting Flange

The probe mounting flange (Vaisala order code: 226061) is a general

purpose mounting flange for 12 mm diameter probes. It can be used to

hold the HMP110 probe in a through-wall installation.

NOTE

The coaxial silicone plug that is delivered with the flange is not suitable

for use with the probe cable of the HMT120.

0911-109

Figure 9

Optional Probe Mounting Flange