Other Commands

Set/show Displayed Quantities

After you have defined the quantities with the CALCS command, use the

DSEL command to show or set the displayed quantities. Number of

parameters determine the display mode (ie. one or two rows). If one

quantity is given, display will switch to one row mode. If two quantities

are given, two-row mode is used. Valid quantities are: RH, T, TD, TDF,

A, X, H, TW, PWS, and PW.

DSEL [q1 q2]

where
Q1 q2

= Quantities to be shown on transmitter display

Syntax for single-parameter transmitters:

DSEL [q]

where
Q

= Quantity to be shown on transmitter display

Examples:

>dsel

1. quantity : RH

2. quantity : T

>

>dsel rh t

1. quantity : RH

2. quantity : T

>

>dsel t

1. quantity : T

>