Calibration Commands

Serial commands that are related to calibration are described below. For

the actual calibration procedures, see section Calibration and Adjustment

on page 50.

Calibrate Humidity Measurement

Use the CRH command to perform a one-point or two-point humidity

(RH) calibration.

CRH

When performing a one-point calibration, you need to place the probe in

a single humidity reference with RH < 50%. Run the command and enter

the exact RH of the reference after the measurement has stabilized. Exit

the command by pressing the ESC key before proceeding to the second

point (see example below). To update the value, press ENTER without

inputting a value.

For two-point calibration, the first point requires a RH < 35% humidity

reference, the second point must be RH > 50%.

Example 1: One-point calibration

>crh

RH: 11.29915720 Ref1 ? 11.3

Press ENTER to continue or ESC to exit

OK

>

Example 2: Two-point calibration

>crh

RH: 11.29915720 Ref1 ? 11.3

Press ENTER to continue or ESC to exit

RH: 75.04306440 Ref2 ? 75

OK

>

Example 3: No calibration, only updating the value

>crh

RH: 20.28000200 Ref1 ?

RH: 20.14000000 Ref1 ?

Calibration terminated with ESC

>