CHAPTER 6
TROUBLESHOOTING
This chapter describes error messages and analog output error behavior,
troubleshooting for possible problems, and provides contact information
for technical support.
Error Codes
HMT120 software includes number of different self diagnostics, such as
flash and program memory checksum, probe communication status,
probe checksum, operation voltage check, oscillator fault check etc.
At startup software checks factory/user settings checksum, program
memory checksum and oscillator fault status. Other checks are made
during runtime. Available error codes are shown in the table below.
Table 7
Error Codes and Texts
Code
Description
Error Text
1
Probe T measurement error
Probe T meas
2
Probe RH measurement error
Probe RA meas
3
Probe communication error
Probe communication
4
Probe checksum error
Probe checksum
5
Probe message form error
Probe message form
6
Program's flash checksum error
Program code checksum
7
Current settings checksum error (RAM)
Settings checksum
8
Factory flash not initialised
Factory defaults empty
9
User flash not initialised
User defaults empty
10
Voltage is too low to operate correctly
Voltage too low
11
Measurements not available
Measurements not available
12
Oscillator fault bit active
HW fault 1
13
Analog output quantity invalid
Analog output quantity invalid
14
Display quantity invalid
Display quantity invalid
Error text is shown by using the ERRS command via the serial interface.
In case of using a HMT120 transmitter with display option the error code
is shown at the display in format "ERR:Code-1[.Code-2][.Code-n]".
Multiple errors are separated with a dot between the error codes.