Chapter 4 _______________________________________________________________ Service Port

VAISALA ________________________________________________________________________ 31

Show System Information

SYSTEM<cr>

Example:

>system

Device Name : HMT120

SW Name : HMT120/130

SW model : HMT120

SW version : 1.0.0.500

Serial number : A1234567

>

Show Firmware Version

VERS

Example:

>vers

HMT120 / 0.1.0.103

>

Serial Line Output Commands

Start Continuous Outputting

Use the R command to start the continuous outputting of measurement

values as an ASCII text string to the serial line. The output always

includes the readings of the currently selected analog output quantities.

R [X]

Example:

>r

RH= 25.10% T= 24.77'C

RH= 25.12% T= 24.96'C

Outputting the results continues in intervals issued with the command

INTV. You can stop the output by entering the S command.

Output data formatting can be defined with FORM command. With X

parameter the transmitter will output RH + T value in predefined format

regardless of what quantities are selected.