Solving typical problems, Table 8, Troubleshooting table – Vaisala HMT120 User Manual

Page 70

background image

User's Guide _______________________________________________________________________

68 ___________________________________________________________________ M211244EN-B

Solving Typical Problems

You can check the error message via the serial interface by using the

ERRS command; see section Error Codes on page 67. In case of

constant error, please contact Vaisala Technical Support.

Table 8

Troubleshooting Table

Problem or Message

Likely Causes and Solutions

Measurement not working, any of

the following errors is active:

- Probe RH measurement error

- Probe T measurement error

- Sensor(s) damaged or missing. Open the

probe filter and check.

- Calibration is not done. Calibrate the probe.

- Check the supply voltage of the transmitter.

Transmitter cannot communicate

with the probe, any of the

following errors is active:

- Probe communication error

- Probe checksum error

- Probe message form error

- Measurements not available

- Check the attachment of the probe to the

transmitter.

- In case of remote probe, check also the

interconnection cables.

- Replace the probe if necessary.

Serial line command not working,

outputs "Unknown command."

- Mistyped or unknown command, check the

command syntax and parameters.

Cannot connect to serial line,

current serial settings of the

module are unknown.

Perform the following steps to connect:

1. Set your terminal settings to 19200 8 N 1

(the default settings of the transmitter).

2. Connect to the service port of the

transmitter.

2. Power up the transmitter, and type "Z" on

the serial line at least five times.

4. The transmitter will come online with the

default settings.

5. Use the

SERI command to set the desired

serial settings, and use the

SAVE

command to save them.

Any of the following errors is

active:

- Current settings checksum error

- Default settings checksum error

Internal error. Perform the following steps:

1. Reset or power cycle the module. Check if

the error disappears.

2. Return the module to factory settings using

the

FRESTORE command. Check again.

3. If the error is still active, contact a Vaisala

service center.

Any of the following errors is

active:

- Program checksum error

- Factory flash defaults checksum

error

- Factory flash not initialized

- Oscillator fault bit active

Faulty transmitter, contact a Vaisala service

center.

Any of the following errors is

active:

- Analog output quantity invalid

- Display quantity invalid

In case of "Analog output quantity invalid"

error select the correct output quantities (set

with the

CALCS command) with ASEL

command and in case of "Display quantity

invalid" error select the correct display

quantities with

DSEL command.