Page 52: Adjustment buttons

background image

User's Guide _______________________________________________________________________

50 ___________________________________________________________________ M211244EN-B

Calibration and Adjustment

Calibrate and adjust the HMT120 using the push-buttons on the

motherboard, serial line commands (see section Calibration Commands

on page 33), or with portable humidity meters HM70 or HMI41.

A calibrator kit is needed for calibration against saturated salt solutions.

The HMK15 Humidity Calibrator and pre-measured certified salts are

available from Vaisala. For further information, please contact your

Vaisala representative.

Vaisala Service Centers also offer accredited calibrations for humidity

and temperature. See contact information on page 69.

It is also possible to remove the HMP110 probe and replace it with a new

one. The old probe can be adjusted using another HMT120 transmitter

body, if you have one available.

NOTE

RH Field Calibration and Adjustment, as instructed in the following

pages, is restricted to a certain adjustment range to minimize the effects

of special circumstances, such as chemical contamination. In case the

probe is outside its adjustment/trim range due to, for example, chemical

exposure, the calibration cannot be completed. In this case, it is

recommended that you change the probe or contact Vaisala Services.

HMT120 Push-Button Calibration

1011-155

Figure 16

Adjustment Buttons

Calibration and adjustment of the HMT120 can be done by using the

adjustment buttons found at the transmitter. The two-point humidity

adjustment is carried out by using two relative humidity references: for

example saturated salt points 11 %RH (LiCl) and 75 %RH (NaCl). The

difference between the two humidity reference points must be at least