Set/Show Available Quantities

The CALCS command shows the abbreviations of the quantities

measured by HMT120. The command can also be used to select the

desired quantities for measurement. Only after the quantities have been

set with the CALCS command, can these quantities be selected for

different outputs (see commands DSEL, ASEL, and FORM). The

quantities in DSEL, ASEL, and FORM must match with CALCS.

CALCS [q1 q2]

where
q1, q2 = Any two of the allowed quantity parameters, i.e. RH, T,

TD, TDF, A, X, H, TW, PWS, PW

Examples:

>calcs

RH Td

>

calcs td t

>

NOTE

Single-parameter transmitters HMT120H and HMT120T have only one

quantity (T for HMT120T, and RH, TD, TDF, A, X, H, TW, PWS, PW

for HMT120H.)

Set/Show Calibration Date

Use the CDATE command to show or set the calibration date field.

(Calibration refers to the calibration of the HMT120's analog outputs.)

CDATE [date]

where
date = Calibration date in format yyyy-mm-dd

Examples:

>cdate

Calibrated : "NOT SET"

>

>cdate 2009-06-22

Calibrated : "2009-06-22"

>