Chapter 6 ____________________________________________________________ Troubleshooting

VAISALA ________________________________________________________________________ 69

Analog Output Error Notification

If the transmitter is unable to operate due to an error, the analog outputs

will be set to an error level.

Default output current in error state is 3.6mA. Error current value can be

changed via the serial interface by using the AERR command; see

section Set/show Analog Output Error Levels on page 38.