Chapter 5 _______________________________________________________________ Maintenance

VAISALA ________________________________________________________________________ 57

13. Remove the probe from the first reference condition and insert

the probe head into the higher humidity reference condition.

14. Press

READY

when the reading has stabilized in the second

reference condition (may take 30 minutes or more). You can

follow the stabilization from the

GRAPH

display.

15. Enter the correct reference value in the second condition with

the arrow buttons. Press

OK

.

16. Confirm the adjustment by pressing

YES

(by pressing

NO

you

return to adjustment mode display and no changes are made). If

the difference between the two reference conditions is less than

50 %RH, adjustment cannot be done.

17. The adjustment is done. Press

BACK

and

EXIT

to return to the

basic display.

18. Turn off the MI70 and detach the connection cable.

LiCl-NaCl Adjustment

This adjustment is done using relative humidity references 11.3 %RH

(LiCl) and 75.5 %RH (NaCl).

Follow steps 1 to 7 on page 54 and continue as follows:

8.

Remove the filter from the transmitter's probe and insert the

probe head into the LiCl salt chamber.

9.

Press

ADJUST

to continue adjusting.

10. Choose

LiCl-NaCl autom.

from the MI70 adjustment menu and

press

SELECT

. Press

OK

to accept the note telling about

references.

11. Press

READY

when the reading has stabilized in the LiCl salt

chamber (may take 30 minutes or more). You can follow the

stabilization from the

GRAPH

display.

12. Remove the probe from the LiCl salt chamber and insert the

probe head into the NaCl salt chamber.

13. Press

READY

when the reading has stabilized in the NaCl salt

chamber (may take 30 minutes or more). You can follow the

stabilization from the

GRAPH

display.

14. Confirm the adjustment by pressing

YES

(by pressing

NO

you

return to adjustment mode display and no changes are made).

15. The adjustment is done. Press

BACK

and

EXIT

to return to the

basic display.

16. Turn off the MI70 and detach the connection cable.