Temperature Field Check and Adjustment by Using a

Calibrated Reference Probe

Follow steps 1 to 7 on page 54 and continue as follows:

8.

Check that the probes are located in equal conditions and wait

until the readings are stabilized (can take 30 minutes or more). If

you are close to the probes, do not breathe in their direction.

9.

Press

ADJUST

to continue adjusting.

10. Press

To same as T

II/I

, press

SELECT

. (MI70 always recognizes the

port to which the HMP70 series probe is connected).

11. Confirm by pressing

YES

.

NOTE

If the temperature difference between the reference probe and HMT120

is too large, the adjustment cannot be done (HM70 will notify you of

this). The available reserve for T adjustment of HMT120 depends on

the initial temperature calibration of the unit.

12. Adjustment is done. Press

BACK

and

EXIT

to return to the basic

display.