Probe Mounting Clamp

The optional mounting clamp makes it easy to install the probe on the

wall of the measurement environment. The probe can be detached for

calibration simply by loosening the lower screw. You can order a single

clamp (Vaisala order code 225501) or a set of 10 clamps (226067).

Installing the entire probe in the measurement environment prevents heat

conduction to the sensor, and is the recommended installation method.

1001-138

Figure 10

Optional Probe Mounting Clamp

CAUTION

Attaching the probe mounting clamp to a conductive wall material

should be avoided, since the potential galvanic connection to the power

supplies and uncontrolled earth current loops could cause measurement

errors or even damage to the HMT120 transmitter.