Restore Probe T Factory Calibration

Use the CTCLR to restore the probe T factory calibration.

CTCLR

Example:

>ctclr

OK

>

Calibrate Analog Output

Use the ACAL command to calibrate the analog current loop outputs.

ACAL

After you give this command, the CH1 analog output of the transmitter

is set to the low limit (4mA). Measure the output and enter the measured

value. After entering the low limit, you must measure and enter the high

limit (20mA) also. The same procedure is then repeated with CH2.

Calculated coefficients are printed after the measured values are given.

Example:

>acal

Ch1 I1 (mA) ? 5.60

Ch1 I2 (mA) ? 18.40

-1.40562890E+03 a0

1.41171900E+03 a1

OK

Ch2 I1 (mA) ? 5.60

Ch2 I2 (mA) ? 18.40

-1.40562890E+03 a0

1.41171900E+03 a1

OK

>

NOTE

Question mark is displayed after certain stabilization time has expired.

Values should be entered only after question mark is displayed. Values

entered before question mark will be discarded.