Hmi41 as a reference meter – Vaisala HMT120 User Manual
Page 64

User's Guide _______________________________________________________________________
62 ___________________________________________________________________ M211244EN-B
HMI41 as a Reference Meter
The HMI41 humidity probe is used to show the correct humidity value
(be sure that the reference probe has previously been adjusted). Connect
the cable and select the appropriate adjustment method as previously
explained, and continue according to the following instructions.
8.
A display similar to the one below should be visible:
Numbers on the first line indicate the transmitter reading, and the
numbers on the second line indicate the reference probe reading.
9.
Let the readings stabilize (may take 30 minutes or more). If you
prefer, you can change the display to show the difference in the
readings. Press
HOLD
and a display similar to the following appears:
The numbers on the first line indicate how much the transmitter
reading differs from that of the HMI41 reference probe. The
numbers on the second line indicate the reference probe reading.
You can return to the previous display by pressing
HOLD
again.
10. When the readings have stabilized, press
ENTER
to conclude the
adjustment. When
ENTER
is pressed, the transmitter reading is
corrected to the reference probe reading. If the adjustment has been
succesful, the following display appears:
11. The data is now stored in the transmitter memory. If the adjustment
has not been succesful, the following display appears:
In this case, perform the adjustment again. Whether the adjustment
was succesful or not, the HMI41 always returns to display the
selected adjustment mode.
12. After succesfully completing the adjustment, turn off the HMI41
and disconnect the cable.