Stop Continuous Outputting

Use the S command to stop the continuous measurement output:

S

Set/show Output Interval

Use the INTV command to show or set the output interval of the serial

line measurement messages (applies when R command or RUN mode is

used). The shortest output interval is one second. This command has no

effect on the operation of the analog outputs.

INTV [n xxx]

where
n

= Time interval in range 1 ... 255

xxx

= Time unit = "S", "MIN", or "H"

Examples:

>intv

Value : 1

Unit : S

>

>intv 3 min

Value : 3

Unit : MIN

>

Output a Reading Once

Use the SEND command to output a single measurement reading from

the transmitter.

SEND [X]

Example:

>send

RH= 25.12 % T= 24.91 'C

>

With X parameter the transmitter will output RH + T value in predefined

format regardless of what quantities are selected.