Classification

Type of protection against electric shock:

Class II Protection from electric shock is achieved by double

insulation.

Degree of protection against electric shock:

Type BF Equipment providing a particular degree of protection against

electric shock regarding

1) allowable leakage current;

2) reliability of protective earth connection (if present).

3) Not intended for direct cardiac application.

Independent testing for Medical Electrical Equipment Standard:

Tested to be in compliance with,

IEC 60601-1 Medical Electrical Equipment – Part 1: General

Requirements for Safety

Tested to be in compliance with applicable requirements of

the Standard,

CAN/CSA C22.2 No. 60601-1-08 M90 Medical Electrical

Equipment – Part 1: General Requirements for Safety

Protection against potential electromagnetic or other interference between the

equipment and other devices.

Tested to be in compliance with:

EN 60601-1-2 Medical Electrical Equipment, Part 1: General

Requirements for Safety-Collateral Standard: Electrical

Compatibility - Requirements and Tests

RTCA-DO160 Airborne Equipment, Sec. 21, Emission of

Radio Frequency Energy

CISPR 11 / EN 55011 Class B Group 1, “Industrial, Scientific,

and Medical (ISM) Equipment”

FCC Part 15, Subpart B – Class B Unintentional Radiators

Method of cleaning and infection control allowed:

Please refer to “Cleaning, Care, and Proper Maintenance”

section of this FreeStyle 5 Patient Manual.

Degree of safety of application in the presence of flammable anesthetic gases:

Equipment not suited for such application.

Mode of operation:

Continuous duty.