MN167-1 rev. – 02/13

Alarm

Rapid

intermittent:

Beep, beep,

beep…

(Yellow) alarm;

intermittent light

Breathing rate is

exceeding the

capacity of the

FreeStyle 5 unit.

Reduce activity, then if

necessary use another

source of oxygen as

available. Contact

your Equipment

Provider.

Alarm

Rapid

intermittent:

Beep, beep,

beep…

(Red) alarm;

continuous light

General

malfunction of the

FreeStyle unit has

occurred.

Turn off the unit.

Change to another

source of oxygen, and

contact your

Equipment Provider.

Figure 16: Alarms Indicator

Dual display:

Alarm condition (yellow)

General malfunction (red)