How to Respond to FreeStyle 5’s Alarm/Light Indicators
Status
Audible
Alarm
Light
Indicates
Action
Indicator
Brief, continuous
at start–up
(Green) pulse;
continuous light
FreeStyle 5 has
been turned on
You may begin to
operate your
FreeStyle 5 unit.
Indicator
No
(Green) pulse;
intermittent light
FreeStyle 5 is
delivering oxygen
as a pulse flow.
Continue using
FreeStyle 5 normally.
Indicator
No
Level indicator
Battery charge
level.
Charge as indicated.
Alarm
Continuous:
Beep
(Yellow) alarm;
continuous light
No breath detected
by the unit for a
predetermined time
period.
Check the cannula
connection. Ensure
that you are breathing
through your nose. If
the alarm persists,
contact your
Equipment provider.
Alarm
Intermittent:
Beep
25% (Yellow)
alarm BATT;
battery indicator
intermittent light
Battery voltage is
too low to operate
FreeStyle.
Connect the FreeStyle
5 unit into a DC outlet
or an AC outlet
immediately.
Alarm
Intermittent:
Beep, beep
25% (Yellow)
BATT; battery
indicator
intermittent light
Battery shutdown
due to low voltage.
Connect the FreeStyle
5 unit into a DC outlet
or an AC outlet
immediately.