Freestyle 5™ en – AirSep MN167-1 User Manual
Page 233

FreeStyle 5™
EN
AirSep
®
Corporation
EN-37
MN167-1 rev. – 02/13
Problem
Probable Cause
Solution
Rapid intermittent alarm
condition, and the (red)
alarm light illuminates
continuously.
A general malfunction
has occurred.
Change to another source
of oxygen as available,
and contact your
Equipment Provider.
Unit does not start on
battery power.
Unit may be hot or
cold if left outdoors
such as in an
automobile.
Allow the unit to reach
normal operating
temperature, which may
take several minutes if
exposed to temperature
extremes. Temporarily
connect your AC or DC
power supply to the unit’s
power inlet and power
source, as needed, to
reset the unit’s internal
battery.
Delay in recharging
internal battery.
Internal battery
exceeds charging
temperature.
Unit may be operated;
however, charging may
not resume until battery
temperature is reduced.
(See Battery Charger
notes.)
Unit alarms while in
automobile and
connected to the
DC outlet.
No power to the unit if
battery depleted and
DC outlet not
charging.
Disconnect the DC power
supply from the
automobile outlet, restart
the automobile, and then
reconnect the DC power
supply into the automobile
DC outlet to reset the
breaker within DC power
supply.
All other problems.
Change to another source
of oxygen as available,
and contact your
Equipment Provider.