Problem

Probable Cause

Solution

FreeStyle 5 does not

operate when a pulse flow

selection button is

pressed.

Battery is discharged.

Malfunction.

Power the unit through the

DC outlet, or an AC outlet.

Contact your Equipment

Provider, and change to

another source of oxygen

as necessary.

A continuous alarm

sounds and the (yellow)

alarm light illuminates

intermittently.

FreeStyle 5 has not

detected a breath for

15 minutes.

Check the cannula

connection.

Ensure that cannula

tubing is not kinked.

Make sure that the

cannula is positioned

properly and that you are

breathing through your

nose. For mouth-

breathing your clinician

may recommend a chin

strap.

Intermittent alarm

condition, and the (yellow)

light illuminates

intermittently.

Battery requires

charging.

Connect to a DC or an AC

outlet within 2 minutes, or

connect to the optional

AirBelt.

Intermittent alarm

condition, and the (yellow)

light illuminates

intermittently and

FreeStyle 5 shuts down.

Battery voltage is too

low to operate the

FreeStyle 5 unit.

Connect to DC or an AC

outlet immediately.

Intermittent alarm

condition, and the (yellow)

alarm light illuminates

intermittently.

Breathing rate has

exceeded the capacity

of the FreeStyle 5 unit.

Reduce activity, and then

turn unit off and back on

again to reset unit. If

necessary, change to

another source of oxygen

as available and contact

your Equipment Provider.