Guidance and manufacturer’s declaration ± electromagnetic immunity

The FreeStyle 5 is intended for use in the electromagnetic environment specified below. The customer

or the user of the FreeStyle 5 should assure that it is used in such an electromagnetic environment.

IMMUNITY

test

IEC 60601

TEST LEVEL

Compliance

level

Electromagnetic environment ± guidance

Conducted RF

IEC 61000-4-6

Radiated RF

IEC 61000-4-3

3 Vrms

150 kHz to

80 MHz

3 V/m

80 MHz to

2.5 GHz

3 Vrms

3 V/m

Portable and mobile RF communications

equipment should be used no closer to any part

of the FreeStyle 5 including cables, than the

recommended separation distance calculated

from the equation applicable to the frequency of

the transmitter.

Recommended separation distance

D= 1.2 x √P

D= 1.2 x √P from 80MHz to 800MHz

D= 2.3 x √P from 800MHz to 2.5GHz

where P is the maximum output power rating of

the transmitter in watts (W) according to the

transmitter manufacturer and d is the

recommended separation distance in metres

(m).

Field strengths from fixed RF transmitters, as

determined by an electromagnetic site surveya

should be less than the compliance level in each

frequency range. b

Interference may occur in the vicinity of

equipment marked with the following symbol:

NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.

NOTE 2 These guidelines may not apply in all situations . Electromagnetic propagation is affected by

absorption and reflection from structures, objects and people.

a. Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones

and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be

predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF

transmitters, an electromagnetic site survey should be considered. If the measured field strength in

the location in which the FreeStyle 5 is used exceeds the applicable RF compliance level above, the

FreeStyle 5 should be observed to verify normal operation. If abnormal performance is observed,

additional measures may be necessary, such as re-orienting or relocating the FreeStyle 5.

b. Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.