AirSep ML-CONC0056 User Manual
Use etask with sequal® eclipse and equinox, Transportable oxygen concentrators, Windows® version
Use eTASK with SeQual® Eclipse and eQuinox
™
Transportable
Oxygen Concentrators
To better assist in titration knowledge, CAIRE has developed eTASK, an interactive software program and smart phone application. By targeting
consistent FiO2 levels, CAIRE’s eTASK quickly illustrates the conversion between continuous flow and pulse dose therapy with autoSAT. Targeting
FiO2s is critical to effective oxygenation, a challenge for ambulating patients due to their changing respiratory rates. eTASK also calculates
battery durations based on the necessary flow rate and the patient’s breath rate.
IMPORTANT
For use only with the SeQual Eclipse and eQuinox. Not for clinical use with other devices. For use only under the supervision
of a medical professional. Equivalent pulse calculations are for reference only. Pulse dose should not be prescribed without
the patient being evaluated by a physician or therapist to determine its effectiveness. Battery durations are typical operating
times for a new battery. Durations may decrease with age, temperature, and breath rate.
Titrate to Saturate
Variable Entry
Continuous Flow
Equivalent Pulse Dose
Continuous Flow (LPM):
3.0
Breaths Per Minute:
20
VT (mL):
500
Concentrator O2 (%):
92.0
Device Type:
Eclipse
Battery Type:
Flow:
3.0 LPM
Est. Battery Life:
1.30 hrs
Pulse Dose Setting:
56 mL
Pulse Setting Number:
3.5
Est. Battery Life:
3.61 hrs
Using the up/down arrows,
input the individual patient
scenario of continuous flow
rate, BPM, tidal volume, and
desired O
2
concentration.
Using the up/down arrows,
input the CAIRE concentrator
the patient will be using
(Eclipse or eQuinox). If using
an eQuinox, select whether
the 12-cell or 24-cell battery
will be used.
The patient’s necessary
continuous flow rate will be
displayed here, along with the
estimated battery duration.
Note that if the required
continuous flow rate is above
3.0 LPM, this is not available
on the Eclipse or eQuinox and
will be shaded in gray.
Equivalent Eclipse or eQuinox
pulse dose setting in mL
and setting number will be
displayed here along with the
battery run time.
Note the pulse setting
numbers are not LPM
equivalents.
Battery run time on pulse
settings will vary based on
breath rate.
eTASK User Guide
1
4
1
2
3
4
2
3
Windows® Version