To better assist in titration knowledge, CAIRE has developed eTASK, an interactive software program and smart phone application. By targeting
consistent FiO2 levels, CAIRE’s eTASK quickly illustrates the conversion between continuous flow and pulse dose therapy with autoSAT. Targeting
FiO2s is critical to effective oxygenation, a challenge for ambulating patients due to their changing respiratory rates. eTASK also calculates
battery durations based on the necessary flow rate and the patient’s breath rate.

IMPORTANT

For use only with the SeQual Eclipse and eQuinox. Not for clinical use with other devices. For use only under the supervision

of a medical professional. Equivalent pulse calculations are for reference only. Pulse dose should not be prescribed without

the patient being evaluated by a physician or therapist to determine its effectiveness. Battery durations are typical operating

times for a new battery. Durations may decrease with age, temperature, and breath rate.

Titrate to Saturate

Variable Entry

Continuous Flow

Equivalent Pulse Dose

Continuous Flow (LPM):

3.0

Breaths Per Minute:

20

VT (mL):

500

Concentrator O2 (%):

92.0

Device Type:

Eclipse

Battery Type:

Flow:

3.0 LPM

Est. Battery Life:

1.30 hrs

Pulse Dose Setting:

56 mL

Pulse Setting Number:

3.5

Est. Battery Life:

3.61 hrs

Using the up/down arrows,

input the individual patient

scenario of continuous flow

rate, BPM, tidal volume, and

desired O

2

concentration.

Using the up/down arrows,

input the CAIRE concentrator

the patient will be using

(Eclipse or eQuinox). If using

an eQuinox, select whether

the 12-cell or 24-cell battery

will be used.

The patient’s necessary

continuous flow rate will be

displayed here, along with the

estimated battery duration.

Note that if the required

continuous flow rate is above

3.0 LPM, this is not available

on the Eclipse or eQuinox and

will be shaded in gray.

Equivalent Eclipse or eQuinox

pulse dose setting in mL

and setting number will be

displayed here along with the

battery run time.
Note the pulse setting

numbers are not LPM

equivalents.
Battery run time on pulse

settings will vary based on

breath rate.

eTASK User Guide

1

4

1

2

3

4

2

3

Windows® Version