beautypg.com

AirSep MN028-1 User Manual

Airsep focus poc, Reading and recording hours

background image

AirSep Focus POC

Reading and Recording Hours

1.) Insert the Focus data cable (MI347) into an open USB port on your PC.

3.) When the program

starts, you will see the

Focus Data Interface box

appear.

2.) Start the program by

clicking on the Focus

Data Interface icon on

your desktop.

4.) Unscrew the two screws

on the bottom of the Focus

and remove the covers.

5.) Lay the Focus unit

on its back, and then

connect the other end

of the data cable to J9

on the circuit board.

The black wire goes

to the top pin as

shown.

Note: Communication is only possible when the Focus unit is in test mode.

To activate the unit’s test mode:

Turn the unit on.

After the 10 second start up time, turn the unit off & on consecutively 3 times (within 4 seconds).

The unit will then start to auto pulse and is now in test mode.

If unit is not pulsing, turn unit off and repeat all steps.

6.) The Data Cable is now connecting the PC and the Focus. Next plug the Focus unit into

AC power, and then put the unit into test mode.

7.) When the Focus unit enters

Test Mode, the Focus Data

Interface Program on your PC

will display hours in the Unit

Run Time box, and the

Compressor Run Time box

within 2 minutes.

2 screws

8.) You may now record the hours from the Unit Run Time box. When recorded, turn

the Focus off, disconnect AC power, disconnect data cable, and re-assemble.

MN028-1

Rev

-

black wire

on top