AirSep MN028-1 User Manual
Airsep focus poc, Reading and recording hours
AirSep Focus POC
Reading and Recording Hours
1.) Insert the Focus data cable (MI347) into an open USB port on your PC.
3.) When the program
starts, you will see the
Focus Data Interface box
appear.
2.) Start the program by
clicking on the Focus
Data Interface icon on
your desktop.
4.) Unscrew the two screws
on the bottom of the Focus
and remove the covers.
5.) Lay the Focus unit
on its back, and then
connect the other end
of the data cable to J9
on the circuit board.
The black wire goes
to the top pin as
shown.
Note: Communication is only possible when the Focus unit is in test mode.
To activate the unit’s test mode:
•
Turn the unit on.
•
After the 10 second start up time, turn the unit off & on consecutively 3 times (within 4 seconds).
•
The unit will then start to auto pulse and is now in test mode.
•
If unit is not pulsing, turn unit off and repeat all steps.
6.) The Data Cable is now connecting the PC and the Focus. Next plug the Focus unit into
AC power, and then put the unit into test mode.
7.) When the Focus unit enters
Test Mode, the Focus Data
Interface Program on your PC
will display hours in the Unit
Run Time box, and the
Compressor Run Time box
within 2 minutes.
2 screws
8.) You may now record the hours from the Unit Run Time box. When recorded, turn
the Focus off, disconnect AC power, disconnect data cable, and re-assemble.
MN028-1
Rev
-
black wire
on top
PDF created with pdfFactory Pro trial version