Freestyle 5, En-42 airsep – AirSep MN167-1 User Manual
Page 238
![background image](/manuals/401231/238/background.png)
EN
FreeStyle 5™
EN-42
AirSep
®
Corporation
MN167-1 rev. – 02/13
Recommended separation distances between portable and mobile RF
communications equipment and the FreeStyle 5
The FreeStyle 5 is intended for use in an electromagnetic environment in which radiated RF
disturbances are controlled. The customer or the user of the FreeStyle 5 can help prevent
electromagnetic interference by maintaining a minimum distance between portable and mobile RF
communications equipment (transmitters) and the FreeStyle 5 as recommended below, according to
the maximum output power of the communications equipment.
Rated
maximum
output power
of transmitter
W
Separation distance according to frequency of transmitter
(m)
from 150kHz to 80MHz
d= 1.2 x √P
from 80MHz to 800MHz
d= 1.2 x √P
from 800MHz to 2.5GHz
d= 2.3 x √P
0.01
0.12
0.12
0.23
0.1
0.38
0.38
0.73
1
1.2
1.2
2.3
10
3.8
3.8
7.3
100
12
12
23
For transmitters rated at a maximum output power not listed above, the recommended separation
distance d in metres (m) can be estimated using the equation applicable to the frequency of the
transmitter, where P is the maximum output power rating of the transmitter in Watts (W) according to
the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
Guidance and manufacturer’s declaration - electromagnetic emissions
The FreeStyle 5 is intended for use in the electromagnetic environment specified below. The customer
or the user of the FreeStyle 5 should assure that it is used in such an environment.
Emissions test
Compliance
Electromagnetic Environment guidance
RF emissions
CISPR 11
Group 1
The FreeStyle 5 uses RF energy only for its internal
function. Therefore, its RF emissions are very low and
are not likely to cause any interference in nearby
electronic equipment.
RF emissions
CISPR 11
Class B
The Freestyle 5 is suitable for use in all establishments,
including domestic establishments and those directly
connected to the public low-voltage power supply
network that supplies buildings used for domestic
purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Complies