“Specifications continued”

Medical equipment needs special precautions regarding EMC and need to be

installed and put into service according to the EMC information provided in this

section.

Guidance and manufacturer’s declaration ± electromagnetic immunity

The FreeStyle is intended for use in the electromagnetic environment specified below. The customer or the user of the

FreeStyle should assure that it is used in such an environment.

IMMUNITY test

IEC 60601 test

level

Compliance level

Electromagnetic environment ±

guidance

Electrostatic

discharge (ESD)

IEC 61000-4-2

± 6 kV contact

± 8 kV air

± 6 kV contact

± 8 kV air

Floors should be wood, concrete or

ceramic tile. If floors are covered with

synthetic material, the relative humidity

should be at least 30 %.

Electrical fast

transient/burst

IEC 61000-4-4

± 2 kV for power

supply lines

± 1 kV for

input/output lines

± 2 kV for power supply

lines

Not Applicable

Mains power quality should be that

of a typical commercial or hospital

environment.

Surge

IEC 61000-4-5

± 1 kV line to line

± 2 kV line to earth

± 1 kV line to line

± 2 kV line to earth

Mains power quality should be that

of a typical commercial or hospital

environment.

Voltage dips, short

interruptions and

voltage variations on

power.

IEC 61000-4-11

<5 % U

T

(>95 % dip in U

T

)

for 0,5 cycle

40 % U

T

(60 % dip in U

T

)

for 5 cycles

70 % U

T

(>30 % dip in U

T

)

for 25 cycles

<5 % U

T

(>95 % dip in U

T

)

for 5s

<5 % U

T

(>95 % dip in U

T

)

for 0,5 cycle

40 % UT

(60 % dip in U

T

)

for 5 cycles

70 % U

T

(30 % dip in U

T

)

for 25 cycles

<5 % U

T

(>95 % dip in U

T

)

for 5 s

Mains power quality should be that of

a typical commercial or hospital

environment. If the user of the FreeStyle

requires continued operation during

power mains interruptions, it is

recommended that the FreeStyle be

powered from an uninterruptible power

supply (UPS) or a battery.

Power frequency

magnetic field

IEC 61000-4-8

3 A/m

3 A/m

Power frequency magnetic fields should

be at levels characteristic of a typical

location in a typical commercial or

hospital environment.

NOTE U

T

is the a.c. mains voltage prior to application of the test level.