Freestyle – AirSep MN131-1 B User Manual

Page 252

background image

EN

FreeStyle™

EN-40

AirSep

®

Corporation

MN131-1 rev. B - 02/13

Problem

Probable Cause

Solution

Intermittent alarm

condition, and the (red)

alarm light illuminates

intermittently.

Breathing rate has

exceeded the capacity

of the FreeStyle unit.

Reduce activity, and then

turn unit off and back on

again to reset unit. If

necessary, change to

another source of oxygen

as available and contact

your Equipment Provider.

Intermittent alarm

condition, and the (red)

alarm light illuminates

continuously.

A general malfunction

has occurred.

Change to another source

of oxygen as available,

and contact your

Equipment Provider.

Unit does not start on

battery power.

Unit may be hot or cold

if left outdoors such as

in an automobile.

Allow the unit to reach

normal operating

temperature, which may

take several minutes if

exposed to temperature

extremes. Temporarily

connect your Universal

AC/DC

power supply to

the unit’s power inlet and

power source, as needed,

to reset the unit’s internal

battery.

Delay in recharging

internal battery.

Internal battery exceeds

charging temperature.

Unit may be operated;

however, charging may

not resume until battery

temperature is reduced.

(See page 9 notes.)