Freestyle™ en – AirSep MN131-1 B User Manual

Page 251

background image

FreeStyle™

EN

AirSep

®

Corporation

EN-39

MN131-1 rev. B - 02/13

Problem

Probable Cause

Solution

FreeStyle does not

operate when a pulse

flow selection button is

pressed.

Battery is discharged.

Malfunction.

Power the unit through the

DC outlet, or an AC outlet.

Contact your Equipment

Provider, and change to

another source of oxygen

as necessary.

A continuous alarm

sounds and the (red)

alarm light illuminates

continuously.

FreeStyle has not

detected a breath for

15 minutes.

Check the cannula

connection.

Ensure that cannula

tubing is not kinked.

Make sure that the

cannula is positioned

properly and that you are

breathing through your

nose. For mouth-breathing

your clinician may

recommend a chin strap.

Intermittent alarm

condition, and the

(yellow) BATT light

illuminates intermittently.

Battery requires

charging.

Connect to a DC or an AC

outlet within 2 minutes, or

connect to the optional

AirBelt.

Rapid alarm condition,

and the (yellow) BATT

light illuminates

intermittently and

FreeStyle shuts down.

Battery voltage is too

low to operate the

FreeStyle unit.

Connect to DC or an AC

outlet immediately.