Problem
Probable Cause
Solution
FreeStyle does not
operate when a pulse
flow selection button is
pressed.
Battery is discharged.
Malfunction.
Power the unit through the
DC outlet, or an AC outlet.
Contact your Equipment
Provider, and change to
another source of oxygen
as necessary.
A continuous alarm
sounds and the (red)
alarm light illuminates
continuously.
FreeStyle has not
detected a breath for
15 minutes.
Check the cannula
connection.
Ensure that cannula
tubing is not kinked.
Make sure that the
cannula is positioned
properly and that you are
breathing through your
nose. For mouth-breathing
your clinician may
recommend a chin strap.
Intermittent alarm
condition, and the
(yellow) BATT light
illuminates intermittently.
Battery requires
charging.
Connect to a DC or an AC
outlet within 2 minutes, or
connect to the optional
AirBelt.
Rapid alarm condition,
and the (yellow) BATT
light illuminates
intermittently and
FreeStyle shuts down.
Battery voltage is too
low to operate the
FreeStyle unit.
Connect to DC or an AC
outlet immediately.