3.

Optional AirBelt (External Battery Belt): FreeStyle can also be

powered by an external battery belt. This belt can be worn around the

waist, and when used in combination with a fully charged internal

battery, supplies power to FreeStyle for up to 10 hours. The belt pack

connects to the FreeStyle unit’s power inlet, and it can be recharged by

connecting it to the battery belt’s own AC power supply.

When using the AirBelt with a fully charged internal battery, the

FreeStyle’s internal battery will deplete before the AirBelt. The AirBelt

must be connected to the FreeStyle unit before the FreeStyle’s internal

battery is depleted. Observe and connect the AirBelt to the FreeStyle

unit before its internal battery discharges to 25% or less.

If the internal battery fully depletes and the FreeStyle unit shuts

down, the unit cannot be restarted with the AirBelt. Should this

occur, connect your FreeStyle to its Universal AC/DC power

supply for a short period of time in order to provide sufficient

internal battery power to start the unit. AirBelt can then be

connected to provide additional run time.

FreeStyle™ Universal AC/DC Power Supply

(Instructions for use)

Connections for removable AC

or DC power cord

Connects to FreeStyle unit

Figure 13

AirSep’s new universal power supply for the FreeStyle Portable Oxygen

Concentrator (POC) is an all-in-one way to keep you fully powered, yet traveling

even lighter – wherever you go. The universal power supply (Part Number

PW022), described in this insert, enables you to operate your unit while charging

its internal battery – all from one component – whether you are driving to the