Freestyle – AirSep MN131-1 B User Manual
Page 234
EN
FreeStyle™
EN-22
AirSep
®
Corporation
MN131-1 rev. B - 02/13
3.
Optional AirBelt (External Battery Belt): FreeStyle can also be
powered by an external battery belt. This belt can be worn around the
waist, and when used in combination with a fully charged internal
battery, supplies power to FreeStyle for up to 10 hours. The belt pack
connects to the FreeStyle unit’s power inlet, and it can be recharged by
connecting it to the battery belt’s own AC power supply.
When using the AirBelt with a fully charged internal battery, the
FreeStyle’s internal battery will deplete before the AirBelt. The AirBelt
must be connected to the FreeStyle unit before the FreeStyle’s internal
battery is depleted. Observe and connect the AirBelt to the FreeStyle
unit before its internal battery discharges to 25% or less.
If the internal battery fully depletes and the FreeStyle unit shuts
down, the unit cannot be restarted with the AirBelt. Should this
occur, connect your FreeStyle to its Universal AC/DC power
supply for a short period of time in order to provide sufficient
internal battery power to start the unit. AirBelt can then be
connected to provide additional run time.
FreeStyle™ Universal AC/DC Power Supply
(Instructions for use)
Connections for removable AC
or DC power cord
Connects to FreeStyle unit
Figure 13
AirSep’s new universal power supply for the FreeStyle Portable Oxygen
Concentrator (POC) is an all-in-one way to keep you fully powered, yet traveling
even lighter – wherever you go. The universal power supply (Part Number
PW022), described in this insert, enables you to operate your unit while charging
its internal battery – all from one component – whether you are driving to the