MN131-1 rev. B - 02/13
o General malfunction
If FreeStyle has a general malfunction, a rapid alarm sounds every ½
second, and the alarm light illuminates red continuously. When this
occurs, the concentration of oxygen that FreeStyle is supplying is below
unit specifications. You should change to another source of oxygen as
available, and contact your Equipment Provider.
This unit is not to be used for life support. Geriatric, pediatric, or
any other patients unable to communicate discomfort while using
this unit may require additional monitoring. Patients with hearing
and/or sight impairment(s) may need assistance with monitoring
alarms.
How to Respond to FreeStyle’s Alarm/Light Indicators
Status
Audible
Alarm
Light
Indicates
Action
Indicator
Brief, continuous
at start–up
(Green) pulse;
continuous light
FreeStyle has been
turned on
You may begin to
operate your FreeStyle
unit.
Indicator
No
(Green) pulse;
intermittent light
FreeStyle is
delivering oxygen as
a pulse flow.
Continue using
FreeStyle normally.
Indicator
No
Level indicator Battery charge level. Charge as indicated.
Indicator
No
(Yellow)
continuous light
Service required.
Return unit to Equipment
Providerfor
inspection/service.